Tool name and version: dccheck 1.10.C.c_0.17
Validation Date: 04-Apr-2012 12:32:53
File name: D:\Documents and Settings\HOEHNHE\Desktop\XMLTest\missingModule.dcm
Number of errors: 9
Number of warnings: 6
Number of checks to be performed manually: 19
| Error type | Module |
| Module ID | Module_Patient |
| Name | Patient |
| Finding | Mandatory Module Patient is missing. |
| References to standard |
|
| Error type | Element |
| Tag | (0010,0010) |
| Name | Patient's Name |
| Finding | Type 2 element (0x0010,0x0010) is missing |
| References to standard |
|
| Error type | Element |
| Tag | (0010,0020) |
| Name | Patient ID |
| Finding | Type 2 element (0x0010,0x0020) is missing |
| References to standard |
|
| Error type | Element |
| Tag | (0010,0030) |
| Name | Patient's Birth Date |
| Finding | Type 2 element (0x0010,0x0030) is missing |
| References to standard |
|
| Error type | Element |
| Tag | (0010,0040) |
| Name | Patient's Sex |
| Finding | Type 2 element (0x0010,0x0040) is missing |
| References to standard |
|
| Error type | Element |
| Tag | ((0018,A001,2),(0040,A170,1),(0008,0104)) |
| Name | Code Meaning |
| Value | Acquisition Equipment
|
| Finding | This Code Meaning is not defined in the Context Groups for this Code Value in this Coding Scheme |
| References to standard |
|
| Error type | Element |
| Tag | ((0018,A001,2),(0040,A170,1),(0008,0104)) |
| Name | Code Meaning |
| Value | Acquisition Equipment
|
| Finding | This Code Meaning is not defined in DCMR Mapping Resource for this Code Value in this Coding Scheme. Expected "Processing Equipment" |
| Error type | Element |
| Tag | ((0018,A001,3),(0040,A170,1),(0008,0104)) |
| Name | Code Meaning |
| Value | Acquisition Equipment
|
| Finding | This Code Meaning is not defined in the Context Groups for this Code Value in this Coding Scheme |
| References to standard |
|
| Error type | Element |
| Tag | ((0018,A001,3),(0040,A170,1),(0008,0104)) |
| Name | Code Meaning |
| Value | Acquisition Equipment
|
| Finding | This Code Meaning is not defined in DCMR Mapping Resource for this Code Value in this Coding Scheme. Expected "Processing Equipment" |
| Warning type | Element |
| Tag | (0018,1000) |
| Name | Device Serial Number |
| Value |
|
| Finding | Type 3 element (0x0018,0x1000) is empty |
| References to standard |
|
| Warning type | Element |
| Tag | (0008,0005) |
| Name | Specific Character Set |
| Value | ISO_IR 100
|
| Finding | Element (0x0008,0x0005) is present and NO extended Chars are used in the message |
| References to standard |
|
| Warning type | Element |
| Tag | ((0018,A001,2),(0018,1000)) |
| Name | Device Serial Number |
| Value |
|
| Finding | Type 3 element (0x0018,0x1000) is empty |
| References to standard |
|
| Warning type | Element |
| Tag | ((0018,A001,3),(0018,1000)) |
| Name | Device Serial Number |
| Value |
|
| Finding | Type 3 element (0x0018,0x1000) is empty |
| References to standard |
|
| Warning type | Element |
| Tag | (0009,0010) |
| Name | Private Creator |
| Value | GEMS_IDEN_01
|
| Finding | Private Creator has been defined, but no Private Element has been assigned in the reserved Block |
| Warning type | Element |
| Tag | (0029,0010) |
| Name | Private Creator |
| Value | GEMS_IMPS_01
|
| Finding | Private Creator has been defined, but no Private Element has been assigned in the reserved Block |