ORL _ Laboratory order response message to a multiple order related to single specimen
OML
ER7
Note that the overriding of IHE Laboratory Profile constraints is only allowed in
national extensions to this framework
The IHE Laboratory Technical Framework requires that applications support
HL7-recommended value that is | (ASCII 124).
2.15.9.1
The IHE Laboratory Technical Framework requires that applications support
HL7-recommended values ^~\& (ASCII 94, 126, 92, and 38, respectively).
2.15.9.2
The codification of these three components is entirely site-defined. It may be
detailed in the national extensions of this framework
2.15.9.3
2.15.9.4
2.15.9.5
2.15.9.6
2.15.9.7
Retained for backward compatibility only
2.15.9.8
Its content is defined within each transaction-specific section of this document
2.15.9.9
This field contains a number or other identifier that uniquely identifies the
message. Each message should be given a unique identifier by the sending system. The
receiving system will echo this ID back to the sending system in the Message
Acknowledgment segment (MSA).
2.15.8.2
2.15.9.11
2.15.9.12
2.15.9.13
2.15.4.1
IHE uses only the HL7 original acknowledgement mode.
2.15.9.15
IHE uses only the HL7 original acknowledgement mode.
2.15.9.16
2.15.9.17
ASCII: The printable 7-bit ASCII character set. 8859/1: The printable characters
from the ISO 8859/1 Character set used by Western Europe. This character set can still be
used, but 8859/15 should be used by preference. This character set is the
forward-compatible version of 8859/1 and includes new characters such as the Euro currency
symbol. ISO IR87: Code for the Japanese Graphic Character set for information interchange
(JIS X 0208-1990).
2.15.9.18
2.15.9.19
2.15.9.20
Note that the overriding of IHE Laboratory Profile constraints is only allowed in
national extensions to this framework
2.15.9.21
The IHE Laboratory Technical Framework authorizes only one of the three values
below, taken from HL7 table 0008
2.15.8.1
This field contains the message control ID from the MSH-10 - Message Control ID of
the incoming message for which the acknowledgement is sent
2.15.8.2
2.15.8.3
2.15.8.4
2.15.8.5
2.15.8.6
2.15.5.1
2.15.5.2
2.15.5.3
2.15.5.4
2.15.5.5
2.15.5.6
2.15.5.7
2.15.5.8
2.15.5.9
2.15.5.10
2.15.5.11
2.15.5.12
Format: [NNN] [(999)]999-9999 [X99999] [B99999] [C any text]
2.15.12.1
2.15.12.2
2.15.12.3
2.15.12.4
2.15.12.5
2.15.12.6
Retained for backward compatibility only
2.15.10.1
2.15.10.2
This field contains the text of the comment. This text may be formatted. In order
to delete an existing comment, the field shall contain empty quotation marks: " " .
Comment text of identical type and source shall be included in the same occurrence of an
NTE segment, and not be split over multiple segments.
2.15.10.3
The IHE Laboratory Technical Framework populates this field with one of the values
in table IHE0364
2.15.10.4
The specific usage of the fields,with usage " Optional" in the table PID, is
explained in the national extensions.
3.4.2.1
3.4.2.2
*
3.4.2.3
3.4.2.4
3.4.2.5
Retained for backward compatibility only
Retained for backward compatibility only
3.4.2.6
Retained for backward compatibility only
Retained for backward compatibility only
If the exact date of birth is not known, the second component of this field can
be used to describe the degree of precision of the information entered in the first
component.
3.4.2.7
Retained for backward compatibility only
3.4.2.8
3.4.2.9
Retained for backward compatibility only
Retained for backward compatibility only
3.4.2.10
3.4.2.11
deprecated as of v2.5
Retained for backward compatibility only
Retained for backward compatibility only
3.4.2.12
3.4.2.13
Format: [NNN] [(999)]999-9999 [X99999] [B99999] [C any text]
3.4.2.14
Format: [NNN] [(999)]999-9999 [X99999] [B99999] [C any text]
3.4.2.15
3.4.2.16
3.4.2.17
the Patient Account Number is not used in some countries: in the US, it must be
valued and the PV1-19 visit number is required when PID-18 identifies an account that
spans more than one Encounter or Visit. In France, PID-18 may be used when a patient
may have multiple visits (each visit having independent transfer and discharge) but
where all the visit need to be linked under one case/episode number.
3.4.2.18
3.4.2.19
3.4.2.20
3.4.2.21
3.4.2.22
3.4.2.23
3.4.2.24
3.4.2.25
3.4.2.26
3.4.2.27
3.4.2.28
3.4.2.29
Retained for backward compatibility only
3.4.2.30
3.4.2.31
3.4.2.32
3.4.2.33
Retained for backward compatibility only
3.4.2.34
3.4.2.35
3.4.2.36
3.4.2.37
3.4.2.38
3.4.2.39
7.4.3.1
This field must be populated in OML messages of transaction T1, in the
context of the use case "Externally placed order with identified specimens" defined
in volume 1. This field is required in OML messages of the T2 transactions. It may
also be used in transactions T3, T4 and T5
7.4.3.2
This version of the IHE of Laboratory Technical Framework doesn't admit
pooling of specimens, therefore the maximum cardinality for the parent specimen is
1. See the data model in volume 1 of this document.
7.4.3.3
7.4.3.4
7.4.3.5
7.4.3.6
7.4.3.7
7.4.3.8
7.4.3.9
7.4.3.10
7.4.3.11
7.4.3.12
2.16.50
7.4.3.13
7.4.3.14
7.4.3.15
This field must be populated if known in OML messages sent by the Order
Placer, within transaction T1, and OML messages sent by the Order Filler within
transactions T2 and T4
7.4.3.16
The date and time when the specimen was acquired from the source. The use of
the Date Range data type allows for description of specimens collected over a period
of time, for example, 24-hour urine collection. For specimens collected at a point
in time, only the first component (start date/time) will be populated
7.4.3.17
Retained for backward compatibility only
Retained for backward compatibility only
4.5.3.14
Retained for backward compatibility only
7.4.3.19
Retained for backward compatibility only
This field is pointless in messages sent by the Order Placer.
7.4.3.20
7.4.3.21
7.4.3.22
7.4.3.23
7.4.3.24
7.4.3.25
2.16.50
For sample receipt verification purposes; may be different from the total
number of specimens that accompany the order.
7.4.3.26
7.4.3.27
7.4.3.28
7.4.3.29
Since the OBX-18 field should not be used; this OBX-19 field is pointless.
This field contains the sequence number of the OBX.
7.4.2.1
7.4.2.2
The usage of LOINC(r) test codes for the identification of tests is strongly
recommended. Details of this free vocabulary can be found at http://www.loinc.org
7.4.2.3
7.4.2.4
7.4.2.5
7.4.2.6
required if available. This field should be valued as described in HL7 V2.5
for all observations for which it is relevant. The References range that figures in
this field is supposed to be related to age and sex of the patient or to other
parameters such as number of weeks of pregnancy when applicable, which makes the
OBX-10 field (nature of abnormal test) unnecessary.
7.4.2.7
required if available. This field is required when applicable. Among the
possible values listed for this field in HL7 table 0078, the actors of IHE
Laboratory Technical Framework should support the values of table 0078 As described
in the above table, the S, R, or I values must be used to indicate the interpreted
result of susceptibilities in microbiology, in case the value field (OBX-5) contains
a numeric value that represents the MIC (Minimum Inhibitive Concentration). In case
the order filler only reports the interpreted result for susceptibilities, the S, R,
or I value could be filled in the value field (OBX-5) with a Value Type (OBX-2) set
to "ST".
7.4.2.8
7.4.2.9
7.4.2.10
This field should be filled according to HL7 Table 0085 described in Chapter
7 of HL7. In this version of the Laboratory Technical Version, the possible values
for this field are a subset of this table:
7.4.2.11
Since the Reference range given by the order filler is the one that applies
to the Observation result taking into account some other parameters like patient age
or sex; there is not need to manage the present field. The Order Result Tracker
should store the observation result and its associated reference range. In case the
Order Result Tracker offers the capability to compare results of the same
observation over different requests, it should then associate each result with its
associated reference range.
7.4.2.12
Retained for backward compatibility only
7.4.2.13
Required if available This field should be valued when the OBX-5 field (Value
field) is also valued. In very exceptional case this information may be unknown by
the Order Filler (case the test is transmitted in a reference lab and observation
date and time is not returned together with the result), this field can be null and
the Order Result Tracker should not generate an error.
7.4.2.14
Retained for backward compatibility only
Required if available This field is required in case the observation was not
produced by the sending organization
7.4.2.15
This field is required when the observation result status (OBX-11) is valued
with "D" or "R" or "P" or "F" or "C" or "X" and the Producer's ID field is not
valued. It should contain the identity of the observer that causes the change of the
observation result status. Only the first component (ID number) of this field is
necessary, provided that it is possible to retrieve the full identity of responsible
person in the Order Filler system with this only ID number.
7.4.2.16
Retained for backward compatibility only
Retained for backward compatibility only
7.4.2.17
Although this information should be available at the Order Filler level, this
field should not be used in this first version of the Laboratory Technical Framework
7.4.2.18
Since the OBX-18 field should not be used; this OBX-19 field is pointless.
7.4.2.19
Retained for backward compatibility only
13.4.3.1
4.5.6.1
13.4.3.3
13.4.3.4
13.4.3.5
4.5.3.15
13.4.3.7
Retained for backward compatibility only
13.4.3.8
13.4.3.9
13.4.3.10
13.4.3.11
13.4.3.12
13.4.3.13
13.4.3.14
13.4.3.15
13.4.3.16
13.4.3.17
13.4.3.18
13.4.3.19
13.4.3.20
8.8.11.4
13.4.3.22
13.4.3.23
13.4.3.24
13.4.3.25
13.4.3.26
8.8.11.7
13.4.3.28
13.4.3.29
13.4.3.30
13.4.3.31
13.4.3.32
13.4.3.33
13.4.3.34
13.4.3.35
13.4.3.36
13.4.3.37
13.4.3.38
13.4.3.39
13.4.3.40
13.4.3.41
13.4.3.42
13.4.3.43
13.4.3.44
This field contains the date/time when the laboratory results are expected to
be available.
This field may be considered the "trigger event" identifier for orders.
Many order control codes are defined in the HL7 table 0119 - Order Control Codes.
The IHE Laboratory Technical Framework allows only a subset of it
4.5.1.1
If the field is valued then its value must match the value of the required
field OBR-2.
4.5.1.2
If the field is valued then its value must match the value of the required
field OBR-3.
4.5.1.3
required if known to the sender. The Placer Group Number represents an
identification of a set of closely related orders, i.e. the whole list of
batteries ordered by the placer to the laboratory for one subject.
4.5.1.4
4.5.1.5
4.5.1.6
4.5.1.7
2.16.50
2.16.67
2.16.78
2.16.82
Retained for backward compatibility only
2.16.82
Retained for backward compatibility only
2.16.78
2.16.78
2.16.83
2.16.38
2.16.50
4.5.1.8
This field contains the date and time of the event that initiated the
current transaction as reflected in ORC-1 Order Control Code. This field is not
equivalent to MSH-7 Date and Time of Message that reflects the date/time of the
creation of the physical message. In OML messages "Status changed" sent by the
Order Filler, this field contains the date/time of the last status change of the
order (ORC-5) or one of the ordered batteries (identified in the following OBR).
4.5.1.9
Retained for backward compatibility only
4.5.1.10
Retained for backward compatibility only
Retained for backward compatibility only
4.5.1.11
Retained for backward compatibility only
Retained for backward compatibility only
This field contains the person (physician) who prescribed the order. See
the data model in volume 1.
4.5.1.12
Retained for backward compatibility only
Retained for backward compatibility only
4.5.1.13
If the field is valued then its value has to match the value of the
required field OBR-17.
4.5.1.14
Format: [NNN] [(999)]999-9999 [X99999] [B99999] [C any text]
4.5.1.15
Retained for backward compatibility only
4.5.1.16
4.5.1.17
4.5.1.18
4.5.1.19
Retained for backward compatibility only
Retained for backward compatibility only
4.5.1.20
This field contains the facility (care unit) placing this order. See the
data model in volume 1.
4.5.1.21
4.5.1.22
deprecated as of v2.5
Retained for backward compatibility only
Retained for backward compatibility only
4.5.1.23
Format: [NNN] [(999)]999-9999 [X99999] [B99999] [C any text]
4.5.1.24
deprecated as of v2.5
Retained for backward compatibility only
Retained for backward compatibility only
4.5.1.25
4.5.1.26
This field contains the date/time when the laboratory results are expected
to be available.
4.5.1.27
Retained for backward compatibility only
4.5.1.28
4.5.1.29
4.5.1.30
This cycle of IHE Laboratory Technical Framework uses doesn't use TQ2
segment, and uses only one occurrence of TQ1 segment, with a single field required
in it : TQ1-9 Priority (CWE).
4.5.4.1
4.5.4.2
2.16.50
4.5.4.3
2.16.16
2.16.38
2.16.50
2.16.50
2.16.50
Constrained to units of time. The codes for unit of measure are
specified in the Unified Code for Units of Measure (UCUM)
[http://aurora.rg.iupui.edu/UCUM].
2.16.39
2.16.38
2.16.38
2.16.50
Constrained to units of time. The codes for unit of measure are
specified in the Unified Code for Units of Measure (UCUM)
[http://aurora.rg.iupui.edu/UCUM].
4.5.4.4
4.5.4.5
2.16.50
4.5.4.6
2.16.50
4.5.4.7
Retained for backward compatibility only
4.5.4.8
Retained for backward compatibility only
Transactions T1, T2 and T3 restrict the usage to the 6 values: S, A, R,
P, C, T of table 0485
4.5.4.9
4.5.4.10
4.5.4.11
4.5.4.12
4.5.4.13
2.16.50
4.5.4.14
4.5.5.1
4.5.5.2
4.5.5.3
4.5.5.4
4.5.5.5
4.5.5.6
4.5.5.7
4.5.5.8
2.16.50
4.5.5.9
4.5.5.10
4.5.3.1
4.5.1.2
4.5.1.3
4.5.3.4
4.5.3.5
4.5.3.6
Retained for backward compatibility only
4.5.3.7
Retained for backward compatibility only
4.5.3.8
Retained for backward compatibility only
4.5.3.9
2.16.50
4.5.3.10
Retained for backward compatibility only
Retained for backward compatibility only
4.5.3.11
4.5.3.12
4.5.3.13
4.5.3.14
Retained for backward compatibility only
4.5.3.15
4.5.1.12
Retained for backward compatibility only
Retained for backward compatibility only
4.5.3.17
Format: [NNN] [(999)]999-9999 [X99999] [B99999] [C any text]
4.5.3.18
4.5.3.19
4.5.3.20
4.5.3.21
4.5.3.22
Retained for backward compatibility only
4.5.3.23
Using ISO
4.5.3.24
4.5.3.25
4.5.3.26
4.5.1.7
2.16.50
2.16.67
2.16.78
2.16.82
Retained for backward compatibility only
2.16.82
Retained for backward compatibility only
2.16.78
2.16.78
2.16.83
2.16.38
2.16.50
4.5.3.28
Retained for backward compatibility only
Retained for backward compatibility only
4.5.3.29
4.5.3.30
4.5.3.31
4.5.3.32
Retained for backward compatibility only
Retained for backward compatibility only
4.5.3.33
Retained for backward compatibility only
Retained for backward compatibility only
4.5.3.34
Retained for backward compatibility only
Retained for backward compatibility only
4.5.3.35
Retained for backward compatibility only
Retained for backward compatibility only
4.5.3.36
Retained for backward compatibility only
4.5.3.37
4.5.3.38
4.5.3.39
4.5.3.40
4.5.3.41
4.5.3.42
4.5.3.43
4.5.3.44
4.5.3.45
4.5.3.46
4.5.3.47
4.5.3.48
4.5.3.49
7.4.3.1
This field must be populated in OML messages of transaction T1, in the
context of the use case "Externally placed order with identified specimens"
defined in volume 1. This field is required in OML messages of the T2
transactions. It may also be used in transactions T3, T4 and T5
7.4.3.2
This version of the IHE of Laboratory Technical Framework doesn't admit
pooling of specimens, therefore the maximum cardinality for the parent
specimen is 1. See the data model in volume 1 of this document.
7.4.3.3
7.4.3.4
7.4.3.5
7.4.3.6
7.4.3.7
7.4.3.8
7.4.3.9
7.4.3.10
7.4.3.11
7.4.3.12
2.16.50
7.4.3.13
7.4.3.14
7.4.3.15
This field must be populated if known in OML messages sent by the Order
Placer, within transaction T1, and OML messages sent by the Order Filler
within transactions T2 and T4
7.4.3.16
The date and time when the specimen was acquired from the source. The
use of the Date Range data type allows for description of specimens collected
over a period of time, for example, 24-hour urine collection. For specimens
collected at a point in time, only the first component (start date/time) will
be populated
7.4.3.17
Retained for backward compatibility only
Retained for backward compatibility only
4.5.3.14
Retained for backward compatibility only
7.4.3.19
Retained for backward compatibility only
This field is pointless in messages sent by the Order Placer.
7.4.3.20
7.4.3.21
7.4.3.22
7.4.3.23
7.4.3.24
7.4.3.25
2.16.50
For sample receipt verification purposes; may be different from the
total number of specimens that accompany the order.
7.4.3.26
7.4.3.27
7.4.3.28
7.4.3.29
13.4.3.1
4.5.6.1
13.4.3.3
13.4.3.4
13.4.3.5
4.5.3.15
13.4.3.7
Retained for backward compatibility only
13.4.3.8
13.4.3.9
13.4.3.10
13.4.3.11
13.4.3.12
13.4.3.13
13.4.3.14
13.4.3.15
13.4.3.16
13.4.3.17
13.4.3.18
13.4.3.19
13.4.3.20
8.8.11.4
13.4.3.22
13.4.3.23
13.4.3.24
13.4.3.25
13.4.3.26
8.8.11.7
13.4.3.28
13.4.3.29
13.4.3.30
13.4.3.31
13.4.3.32
13.4.3.33
13.4.3.34
13.4.3.35
13.4.3.36
13.4.3.37
13.4.3.38
13.4.3.39
13.4.3.40
13.4.3.41
13.4.3.42
13.4.3.43
13.4.3.44